The range of centrifuge is specifically designed for use in Oil Testing for tests as per ASTM and IP standards. The unit is fabricated from steel and epoxy coated.

Parameters

OC2F

Control

Microprocessor with Variable Frequency Drive

Max. Speed

3000 RPM

Max. RCF

1480 x g

Choice of Tube Size

100 ml

Max. Capacity

400 ml

Motor

Brushless Induction Motor

Display

Digital LCD

Program Features

User defined RPM/RCF, Acceleration, Deceleration & Timer with 25 Programs

Acceleration/ Deceleration

12 - 510 Secs

Timer

1 - 120 Min

Brake

Dynamic Brake

Safety

Sturdy Cabinet, Lid Lock, Imbalance Protection, Gas Spring

Security

Password Protection (Optional)

Choice of Rotor Heads (Swing Out)

2

Power Supply

230V, 50 Hz

Swing Out Rotor Head with carriers suitable for use with Centrifuge Model OC 2 F:

Model

Capacity

Type of Tests

RPM

RCF xg

OC 1 P + OC 2

4 Place S/O for 100 ml Pear shaped tubes

ASTM-96

3000

1480

OC 1 T + OC 4

4 Place S/O for 100 ml Taper bottom tubes,sulfonation flask & bob cock bottles

ASTM-D-91, ASTM-D-96,
ASTM-D-1796
ASTM-D-893, IP-75-57,
ASTM-D-483
ASTM-D-484,
ASTM-D-1019, IP-145

3000

1480

Standard Tests:

Type of Tests

Specification

Recom Max. RPM.

mended Max. Time

Water & sediments in crude oil

ASTM-D-96

800

10 Mts.

Water & sediments in petroleum products

IP-75

800

10 Mts.

Water & sediments in fuel oils

ASTM-D-1796

800

10 Mts.

Precipitation number of lubricating oils

ASTM-D-91

700

10 Mts.

Stoddard solvent

ASTM-D-483

600

10 Mts.

Olefinic plus aromatic hydrocarbons

in petrolleum distillates

 

ASTM-D-1019

 

400

 

15 Mts.

Olefinic plus aromatic contents

ASTM-D-893

700

20 Mts.

Unsulfonated residue of petroleum plant

ASTM-D-483

100

10 Mts.

Residue in liquified petroleum (L.P.)

ASTM-D-1266/1657

800

10 Mts.

Trace sediments in lubricating oil

ASTM-D-287/611/86

700

10 Mts.